Ürün Açıklaması
1 Advanced Testing Method
* Inductive dual magnetic circuit method makes use of electro-magnetic fields.No affection of magnetic field attenuation.
* No disruptions from icterus,hemolysis, chyles and turbid particles.
* Test result by sensing relative motion of magnetic beads.No affection of the plasma viscosity.
* Intelligent switching between normal and weak vibration.More accurate result of low fiber sample.
2 Unique Design of Dual Probes
* Independent sample and reagent probes.
* Liquid surface induction inside probes.
* Instantly heating to 37 °C with precise control inside sample probe.
3 High-speed Motion Mechanism
* 1000 continuous cuvettes give no test suspending for cuvettes feeding.
* Synchronic motion mechanical manipulators on X,Y and Z orientations.
* High-speed cuvettes feeding and random assortment of testing channels.No waiting period for pre-heating.
4 Accurate Temperature Control System
* 16°C temperature control ability of reagent positions.
* 37°C thermostatic control of testing and pre-heating positions.
* Inner heating ability of the probe keeps constant temperature of reagent.
5 Easy Operation Process
* Compatible to any tube.
* Random assortment of testing items on all channels.
* Emergency testing on any position.
* Abnormal results prompting and automatic re-test.
* Error prompting and insufficiency alarm for reagent and cleaning fluid.
* Testing report in various formats.
6 Excellent Testing Performance
* Up to 200 tests per hour of single item.
* 30 samples per hour of mixed tests of four rountine items.
* High lineariyt with correlation coefficient > 0.980.
* Wide FIB testing range of 0.3-12g/L.
7 Wonderful System Solution
* Full automated analyzer.
* Original reagents for all items.
* Application services on instrument, reagent and control.
* High correlation and traceability of testing result.
Specification
Testing Method: Inductive dual magnetic circuit method. Signal from eddy current sensor.
Testing Items: PT,APTT,FIB,TT,AT-III,HEP,LMWH,PC,PS and all types of endogenous and exogenous factors.
Throughput: 200 Ts/h.
CV: <3 %
Testing Positions: 4
Pre-heating Positions: 10
Reagent Positions: 6, with cooling and stirring ability.
Sample Positions: 30, interchangeable and extensible sample rack compatible to any tube.
Cuvettes: 1000 units/ load.
Port: USB and RS232
Data Transmission: Dual way, HIS/LIS supported.